- Determination of moisture in liquefied and gaseous samples
- Inlet pressure up to 200 bar/2900 psi
- Determination of pressure in the sample loop
- Automatic pressure regulation
- Transfer line with direct injection
- Automatic rinsing bypass and steps for rinsing
- Measuring cell without diaphragm (only one electrolyte required)
- Setting of application-specific methods
- Sulphur trap eliminating the side reactions caused by sulphides/H2S
- Type of result: µg, ppm (gas volume), Vppm, Mppm, Mol ppm by using the formula generator
Areas of Application
LPG, LNG
- propane, propene, butane, butene, butadiene
- ethylene oxide
- chlorinated hydrocarbons, e. g. methylene chloride, ethylene chloride, vinyl chloride
Analysis of refrigerants:
- halogenated hydrocarbons
Analysis of permanent gases:
- natural gas
- technical gases
- mixtures of test gases
Advantages
- Sulphur removal cartridge eliminating the side reactions caused by sulphides/H2S
- No interference calculation required
- 250 measurements can be performed in 48 hours
- Totally automated process, no operator input required for the test
- Suitable to test all gas types without any calibration or adjustments
- No separate rinsing gas is required
- Rinsing process is fully automated
- No balance is required
- High sample throughput and long reagent life
- Compact device
Measurement method: | Coulometric Karl Fischer titration |
Sample: | Pressurized gas sample (LNG, LPG) |
Sample dosing: | Pressurized bottle or directly from the gas line |
Pressure reducer: | internal (with heating element) |
Sample loop: | 300 mL (gas) |
Rinsing and dosing: | 0 … 15 steps for each, adjustable |
Measuring range: | 1 μg ... 100 % |
Resolution: | 0.1 ppm |
Detection limit: | 1 ppm |
Power supply: | 230 V/50 Hz; 115 V/60 Hz |
Dimensions: | 33 x 49 x 48 cm (W x D x H) |
Weight: | 24.0 kg |
Device control: | PC software (PC not included in the scope of delivery) |